23.3.3 Parameter Descriptions and Format
The text string that is output contains 40 parameters. e.g.
60,2,6,10,23,100,4,2,100,1,0,5,20,0,45,20,30,160,280,0,45,0,15000,60,3,10,145
,139,3,25,99,22,13,0,824,10,3275,6,39,64,
A description of each parameter follows;
23.3.4 List of Parameters
Param. Nº Description Units
Details
1 Set BPM breaths/minute 0 to 150
2 Set CPAP mbar Set CPAP pressure 0 to 35 mbar
3 Set Tidal Volume 0.2 ml Set Target Tidal volume 10 to
1000 (2 to 200ml)
4 Set Insp. Time 0.01 sec. Set insp. time or max insp time,
10 to 300 (0.10 to 3.00 seconds)
5 Set PIP mbar Set PIP pressure 0 to 65 mbar
6 Set FiO2 % Set O2 Concentration 21 to
100%
7 Set HFO Delta P mbar Set HFO Delta P 4 to 180 mbar
8 Set HFO Mean mbar Set mean 0 to 35 mbar
9 Set HFO Rate 0.1 Hz Set HFO rate 30 to 200 ( 3 to
20Hz)
10 Ventilation Mode n/a 30 = Ventilation Off
0 = CPAP
1 = CMV
31 = PTV
32 = PSV
3 = SIMV
4 = HFO Only
5 = HFO + CMV
11 TTV Status n/a 0 = off