Page 225 of 292
24.3 Alarm Descriptions and Actions to be Taken
Alarm 1. Monitor Failure
Alarm message.................................... Monitor failure
Alarm sub message ............................ Remove ventilator from service
Priority of alarm................................... 1
Monitor mode....................................... N/A
Can alarm be muted ........................... No
Sounder Priority .................................. High
Alarm Description:
If the monitor system fails this alarm is generated by the controller system.
Action: Remove patient to alternative form of ventilation, then remove
ventilator from service.
Alarm 2. Sustained Sub-ambient Pressure
Alarm message.................................... Sustained sub ambient pressure
Alarm sub message ............................ Controller operation temporarily
disabled
Priority of alarm................................... 2
Monitor mode....................................... All
Can alarm be muted ........................... No
Sounder Priority .................................. High
Alarm Description:
If the proximal pressure falls below -2mbar for >50ms a sub-ambient alarm is generated
and a message will be sent to the control sub-system to shut off all gasses. If the gas is not
cut off during the next 50ms the monitor will intervene and cut all the gasses. Fresh gas
reinstated after 6 seconds ventilation restarted after a further 2 seconds.
In all HFO modes it is the mean pressure that is checked for sub-ambient level.
Action: Remove patient to alternative form of ventilation, then remove
ventilator from service.