Page 259 of 292
Ensure that the detergent solution does not enter the unit or the exhalation
block gas ports on the side of the machine.
(Touch Screen) Do not use any abrasive cleaners on the touch screen
surface.
Flow Sensor
Wash in hand hot water/mild general purpose detergent solution (as prescribed
by the appropriate hospital authority). Rinse with sterile water.
Warning: Do not clean the flow sensor with compressed air or water jet. As
this will destroy the sensor wires
Exhalation block. The exhalation block can be immersed and agitated in the
detergent solution. Do not insert any objects into the exhalation block. Rinse the
exhalation block in clean water, it must be allowed to dry thoroughly before
sterilization.
27.4 Disinfection Method
Note: Alcohols such as 70% isopropanol have a good activity against
bacteria and viruses. They should only be used after all visible surface dirt
has been removed from the area to be disinfected.
Ventilator. For disinfection use two clean, disposable, absorbent, non-shedding
cloths. Wipe clean with the first cloth using Alcohol (70% isopropanol). Wipe dry
with the remaining cloth.
Flow Sensor
Immersion in disinfectant recommended for use with plastic materials. Immersion
times and concentrations must be in accordance with manufacturer’s instructions.
Rinse with sterile water.
Warning: Disinfectants containing compounds similar to PHENOL or
ALKYLAMINES (Glucorrotamine) are unsuitable.
Exhalation block. The exhalation block can be immersed in Alcohol (70%
isopropanol). The exhalation block must be allowed to dry thoroughly before
sterilization.