28.1.4 PSV
Inspiratory Time: 0.1 to 3.0 seconds, steps of 0.01 seconds
CPAP Pressure: 0 mbar to 20 mbar
Inspiratory Pressure: 0 mbar to 65 mbar
Volume Targeting: 2 ml to 200 ml
FiO
2
: 21% to 100%
Backup breaths are pressure limited, time cycled.
28.1.5 SIMV
BPM: 1 to 150
I:E Ratio: Calculated from BPM and Inspiratory time settings.
(11.2:1 to 1:600)
Inspiratory Time: 0.1 to 3.0 seconds, steps of 0.01 seconds
PEEP Pressure: 0 mbar to 20 mbar
Inspiratory Pressure: 0 mbar to 65 mbar
Volume Targeting: 2 ml to 200 ml
FiO
2
: 21% to 100%
Breaths are flow cycled, pressure limited. Backup breaths are time cycled
pressure limited.
28.2 HFO Ventilation
28.2.1 HFO Only
Frequency Range: 3-20Hz. (1Hz resolution.)
I:E Ratio 1:1
Delta Press Range: 4 mbar to 180 mbar
Mean airway range: 0 mbar to 35 mbar maximum
FiO
2
: 21% to 100%
28.2.2 HFO+CMV
BPM: 1 to 150
Inspiratory Time: 0.1 to 3.0 seconds, steps of 0.01 seconds
Frequency Range: 3-20Hz. (1Hz resolution.)
I:E Ratio Calculated from BPM and Inspiratory time settings.
(11.2:1 to 1:600)
Inspiratory Pressure: 0 mbar to 65 mbar
Delta Press Range: 4 mbar to 180 mbar
Mean airway range: 0 mbar to 35 mbar maximum
FiO
2
: 21% to 100%