22. Preventative Maintenance and Overhaul Schedule
Warning: Preventative Maintenance, Overhaul and Calibration of this
ventilator should only be carried out by an SLE trained hospital engineer or
an SLE service engineer.
Interval Maintenance type Kit type Part Nº
6 months Preventative maintenance Uses 6 month maintenance kit N9010/06
12 months Prevenative maintenance Uses 12 month maintenance kit N9010/12
18 months Prevenative maintenance Uses 6 month kit maintenance N9010/06
24 months Overhaul Uses 24 month overhaul kit N9010/24
30 months Prevenative maintenance Uses 6 month maintenance kit N9010/06
36 months Prevenative maintenance Uses 12 month maintenance kit N9010/12
42 months Prevenative maintenance Uses 6 month maintenance kit N9010/06
48 months Overhaul Uses 48 month overhaul kit N9010/48
A service manual is available for use by qualified engineers who have been
trained by SLE on this product. Contact SLE or your distributor for further
information.