26. Patient Circuits
The ventilator can use both single use and autoclavable patient circuits.
26.1 Warnings for Patient Circuit Use.
Use only SLE approved patient circuits.
Applicable to single use and re-usable patient circuits.
Make sure that all connections are made properly and are tight before use.
For circuit heating, follow the directions provided with the humidifier.
Do not use in applications where gas temp. at outlet of humidifier exceeds
55ºC.
Do not cover the patient circuit with anything which may cause the tube to
overheat.
Do not autoclave water trap if medications containing chlorinated or
aromatic hydrocarbons are used or with the water container fitted.
The circuit must not be allowed to come into contact with the skin - an
unconscious patient is not able to react to any heat or possible burn risk.
Keep water trap upright with cup at bottom. Empty container before water
reaches MAX Level Line. After emptying, ensure the water trap is fitted
correctly. Check the operating pressures are as before emptying the water
trap.