2.5.2 SIMV with TTV (Targeted Tidal Volume)
This is as for basic SIMV with or without Pressure Support, where the inspiratory phase shall
be terminated if the set volume has been delivered.
The User sets the following:-
• Enables Targeted Tidal Volume and selects volume to be delivered
•Max PIP
•Max Ti
2.6 HFO (SLE5000 only)
High Frequency Oscillation
In this mode, the ventilator shall deliver continuous high frequency oscillation. There is no
patient interaction.
The User sets the following:-
•HFO rate
• Mean pressure
• Pressure amplitude (or Delta P)
•Fi0
2
• High and low alarm pressure thresholds
• Tidal volume, high and low alarm thresholds
• Minute volume high and low alarm thresholds
2.7 HFO+CMV (SLE5000 only)
A combination of oscillations during the expiratory or inspiratory & expiratory phase of a time
cycled, pressure limited breath in CMV mode.
The User sets the following:-
•BPM
• Inspiratory Time (Ti)
• Peak Inspiratory Pressure (PIP)
• Positive End Expired Pressure (PEEP)
•HFO rate
• HFO activity (oscillations in inspiratory and expiratory phases or expiratory phase
only)
• Pressure amplitude (Delta P)
•Fi0
2
• High and low alarm pressure thresholds
• Tidal volume, high and low alarm thresholds
• Minute volume high and low alarm thresholds