2.5 SIMV (Synchronized Intermittent Mandatory Ventilation)
The frequency of mandatory breaths is determined by the BPM control. When a mandatory
breath is due an assist window opens and waits for a patient's inspiratory effort. When this
occurs the ventilator delivers a synchronised breath (SIMV breaths). Once the breath has
been delivered the assist window closes until the next set breath is due.
If the ventilator does not see a patient's attempt to breathe before the end of the defined time
window then a mandatory breath is delivered. The mandatory breath point is the Time
Window minus the Inspiratory Time.
User sets the following:-
•BPM
Apnoea time (Only if backup breath rate is 19 breaths per minute or lower)
Positive End Expired Pressure (PEEP)
Peak Inspiratory Pressure (PIP)
Inspiratory Time (Ti)
•Fi0
2
Flow breath detection threshold (Set Trigger) - The patient effort required to satisfy
the apnoea alarm.
High and low pressure alarm thresholds
Tidal volume, high and low alarm thresholds
Minute volume, high and low alarm thresholds
Assist windows
Time windows
60
BPM
Time window =
Inspiratory time
Mandatory breath point