Page 267 of 292
28.3.3 Pressure
Real Time Pressure: Resolution 1mbar
Drift detection: detects drift when there is a pressure difference of more
than 5mbar between the two pressure transducers. Only
active when the measured pressure is below 70mbar.
Sampling Time: 2ms
Peak Pressure: 0 to 175 mbar (resolution 1mbar)
PEEP Pressure: 0 to 175 mbar (resolution 1mbar)
Mean Pressure: -175 to 175 mbar (resolution 1mbar).
Delta P In HFO combined mode Delta P is measured during