28.2.4 Controls
28.3 Measurement
28.3.1 Flow and Volume
Flow Sensor Type: 10 mm dual-hot-wire-anemometer with replaceable hot-
wire subassembly, autoclaveable. Sensor electrically
isolated.
Flow Rate: 0.2 lpm to 32 lpm (Accuracy ±8 % maximum)
Expiratory Tidal Volume:
0 to 999ml (0.1ml)
Expiratory Minute Volume:
0 to 18L (Resolution: 1ml)
Deadspace: 1 ml
Weight: 10 g
Conventional Ventilation and combined modes only:
Tube Leakage: 0 to 50% (Resolution: 5%)
Averaged over 5 breaths
Breath Rate (total): 0 to 250 BPM (Resolution: 1BPM)
Dynamic Compliance: 0 to 100ml/mbar (Resolution: 1ml/mbar)
C20/C: Resolution 0.1
Sampling Time: 2ms
Resistance: 0 to 1000 mbar.second/l
Triggering: Inspiratory flow (0.2 to 10 lpm)
Above values are obtained under ATPD (ambient temperature and pressure, dry)
conditions.
HFO only:
DC0
2
: Resolution 1
28.3.2 Oxygen Concentration
Range: 21% to 100% (Resolution 1%)
Battery power termination: Terminates backup battery power supply