28.8 Power, Dimensions, Standards etc.
Voltage : 100-240V/ 50-60Hz
Power : 115 VA
Fuses : 100-250V~50-60Hz : Fuse T(2.0AH) 250v
Battery Back up 12 volt. Up to 45-60 minutes dependent on mode.
Battery life 10 years
Battery Charging Full charge 24 Hours
80% charge 8 hours
Rating: Class 1. Unit must be earthed.
Operating Environment:
Temp: 10ºC-40ºC
Humidity: 30-75% (Non condensing)
Size, Ventilator only: 330mm W x 330mm H x 470mm D
Height on pole: 1140 mm
Weight (ventilator only): 23.6 Kgs
Constructed to conform to:
BS EN 475: 1995
BS EN 794-1:1997
BS EN 60601-1:1990
BS EN 60601-1-2:1993
BS EN 60601-1-4: 1996
Medical Devices Directive (93/42/EEC)
Patient Circuit Required:
Model: 10 mm Single use for use with incubators (SLE Part Nº: N5188)
Model: 10mm Single use for use cots (SLE Part Nº: N5188/02
Model: 10 mm Re-usable (SLE Part Nº: N2391/50)
28.9 Environmental Storage Conditions
When packed for transport or storage;
Ambient Temperature -40°C to +70°C
Relative Humidity 10% to 90% non condensing
Atmospheric Pressure 500hPa to 1060hPa