12 Set Termination
Sensitivity
% Set percentage
0 to 50 = percentage of max. flow
that triggers termination,
if value < 0 then termination
sens. is off
13 Set Breath Trig.
Threshold
lpm for flow
triggering.
No units for
pressure
triggering.
2 to 200 (0.2 to 20lpm for flow
triggering).
14 Set Waveshape n/a Set Rise time value 0 to 25 (rise
time is proportional to the value)
15 Set Patient Leak
Alarm
% Percentage leak at which alarm
occurs.
10 to 50%.
Values above 50 = alarm off.
16 Set Apnoea Alarm seconds 5 to 60 seconds
17 Set Low Pressure
Alarm
0.1 mbar -1200 to 1100 (-120 to 110 mbar)
18 Set Cycle Fail Alarm 0.1 mbar 0 to 1150 ( 0 to 115 mbar)
19 Set High Pressure
Alarm
0.1 mbar 100 to 1200 (10 to 120 mbar)
20 Set Low Tidal Vol.
Alarm
0.1 ml 0 to 2970 (0 to 297ml)
21 Set High Tidal Vol.
Alarm
0.1 ml 30 to 3000 (3 to 300ml)
22 Set Low Minute Vol.
Alarm
ml 0 to 18000 (0 to 18l)
23 Set High Minute Vol.
Alarm
ml 1000 to 18000 (1 to 18l)
24 Measured Total BPM breaths/min Total breath count over the last
minute 0 to 255.
25 Measured CPAP mbar -175 to +175 mbar
26 Measured Insp. Time 0.01 sec. 0 to 32768 (0 to 327.68 seconds)
27 Measured Insp.
Volume
0.1ml 0 to 32768 (0 to 3.2768 l)