EasyManua.ls Logo

SLE SLE5000 - Page 220

SLE SLE5000
292 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 220 of 292
28 Measured Exp.
Volume
0.1ml 0 to 32768 (0 to 3.2768 l)
29 Measured PEEP mbar -175 to 175 mbar
30 Measured PIP mbar -175 to 175 mbar
31 Measured FiO2 % 0 to 100 %
32 Measured HFO Delta
P
mbar -175 to 175 mbar
33 Measured HFO Mean mbar -175 to 175 mbar
34 Trigger Count breaths/min Number of patient triggers in the
last minute 0 to 255
35 Measured Minute Vol. ml 0 to 32768 ( 0 to 32.768l)
36 Measured Leak % 0 to 100 % leak
37 Measured Resistance 0.1 mbar.sec/l Airway resistance 0 to 32768 (0
to 3276.8 millibar seconds per
litre)
38 Measured Compliance 0.1 ml/mbar Total Compliance 0 to 255 (0 to
25.5 millilitres per millibar)
39 Measured C20/C 0.1 This value is a ratio of the
compliance over the last 20% of
the pressure rise compared to the
total compliance. 0 to 255 (0 to
25.5)
40 Current Alarm n/a This value indicates the currently
displayed alarm condition (see
table opposite)
Param. Nº Description Units
Details

Table of Contents

Related product manuals