Page 130 of 292
Pass Fail
Step 23: Leak alarm.
Alarm triggered and cancelled.
Step 24: Block alarm.
Alarm triggered and cancelled.
Step 25: Mains failure alarm.
Alarm triggered and cancelled.
Step 26: Gas supply alarms.
No Air supply Alarm triggered and cancelled.
Step 27: Gas supply alarms.
No O
2
supply Alarm triggered.
Step 28: Gas supply alarms.
No Gas Alarm triggered and cancelled.
Step 29: Flow sensor disconnect alarm.
Alarm triggered and cancelled.
Step 30: Cycle fail alarm.
Alarm triggered and cancelled.
Step 35: HFO waveform test.
Ventilator is oscillating, waveforms
appear and I:E ratio = 1:1.
Step 37: HFO+CMV functional test.
The ventilator cycles with oscillations in both
the inspiratory and expiratory phases. The
waveforms appear in the waveform windows
the I:E ratio in the breath parameter panel
reads 1.0 : 1.0.
Step 38: HFO+CMV functional test.
The ventilator cycles with oscillations in the
expiratory phase only.
Step 40: Pressure Change detected Alarm.
Alarm triggered and cancelled.
Refer the ventilator to qualified service personnel if it did not pass one of
the standard tests.