With the sensor now occluded press the
Calibrate button (A). This initiates
calibration cycle for the flow sensor.
Keep the sensor occluded until the words
Calibration Complete appears above the Calibrate
button.
The sensor has now been calibrated and can be
fitted into the patient circuit.
Note: The flow sensor should be calibrated every 24 hours whilst in use if
patient’s condition permits.