EasyManua.ls Logo

SLE SLE5000 - Page 185

SLE SLE5000
292 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Page 185 of 292
Prior to first use as well as after each use Clean or Disinfect/Sterilize the sensor.
Cleaning: A soap solution or mild alkaline solution should be used.
Disinfection: Use commercially available disinfectants that are recommended
for use with PLASTIC MATERIALS. Immersion times and
concentrations stated must be in accordance with manufacturer’s
instructions
Note: Disinfectants containing compounds similar to PHENOL or
ALKYLAMINES (Glucorrotamine) are unsuitable.
Note: Eliminate all residues of cleaning agents and disinfectant
used by thoroughly rinsing with sterile water after each
cleaning and disinfecting procedure.
Sterilisation For sterilization a validated steam sterilization according to BS
EN554 can be applied. Autoclave at
134ºC (277ºF) (Allowable variation of temperature of +3ºC) at
220kPa (32psi) with a minimum holding time of 3 minutes.
or
121ºC (248ºF) (Allowable variation of temperature of +3ºC) at
96kPa (14.1psi) with a minimum holding time of 15 minutes.
or
Ethylene Oxide at 55ºC (131ºF) plus the recommended time for
aeration
Note: The sensor must not be connected to other standard
connectors when sterilised to prevent cracking. Ensure no
other components/items are lying on the sensor during the
sterilisation process.
Before each use the sensor must be checked for damage. Damaged parts must
not be used.
Warning: Do not use any nebulized gases (medications, salt solutions, etc.)
in conjunction with the sensor as they are likely to degrade the performance
of the sensor and subsequent displayed accuracies.

Table of Contents

Related product manuals