Low pressure alarm
with CPAP/PEEP/
Mean at zero and
PIP/Delta P at zero.
Alarm Message:
Low Pressure.
Air and Oxygen supply failed. If generated whilst connected to
a patient, remove patient to
alternative form of ventilation.
Check Air and Oxygen supplies/
connections.
No O
2
supply alarm.
Alarm Message:
No O2 Supply.
Oxygen supply not connected to
ventilator.
Pressure has dropped to below
4bar.
Oxygen supply failed.
Check Oxygen supply /
connections.
Ensure that the Pressure is at
4bar
If generated whilst connected to
a patient, remove patient to
alternative form of ventilation.
No Air supply alarm.
Alarm Message:
No Air Supply.
Air supply not connected to
ventilator.
Pressure has dropped to below
4bar
Air supply failed.
Check Air supply / connections.
Ensure that the Pressure is at
4bar
If generated whilst connected to
a patient, remove patient to
alternative form of ventilation.
Battery fault alarm.
Alarm Message:
Battery Fault.
The internal battery has failed or
the power supply has developed
a fault.
Remove ventilator from service.
Note alarm message and refer
ventilator to qualified service
personnel.
Battery Low alarm.
Alarm Message:
Battery Low.
Battery has reached minimum
charge level.
Restore mains power.
If mains power cannot be
restored, remove patient to
alternative form of ventilation.
Note alarm message and refer
ventilator to qualified service
personnel.
Symptom Possible Cause Remedy