Page 210 of 292
Unexpected rise in
mean P alarm.
Alarm Message:
Unexpected rise in
mean P.
The mean pressure has
increased by more than 5mbar.
Check ventilator pressures.
Check the patient circuit.
Press autoset to for new alarm
thresholds.
Unexpected drop in
mean P alarm.
Alarm Message:
Unexpected drop
in mean P.
The mean pressure has
decreased by more than 5mbar.
Check ventilator pressures.
Check the patient circuit and
water trap.
Press autoset to for new alarm
thresholds.
Unexpected rise in
max P alarm.
Alarm Message:
Unexpected rise in
Max P.
The maximum pressure has
increased by more than 5mbar.
Check ventilator pressures.
Check the patient circuit.
Press autoset to for new alarm
thresholds.
Unexpected drop in
max P alarm.
Alarm Message:
Unexpected Drop
in Max P.
The maximum pressure has
decreased by more than 5mbar.
Check ventilator pressures.
Check the patient circuit and
water trap.
Press autoset to for new alarm
thresholds.
Unexpected rise in
min. P alarm.
Alarm Message:
Unexpected rise in
Min P.
The minimum pressure has
increased by more than 5mbar.
Check ventilator pressures.
Check the patient circuit.
Press autoset to for new alarm
thresholds.
Unexpected drop in
min. P alarm.
Alarm Message:
Unexpected Drop
in min P.
The minimum pressure has
decreased by more than 5mbar.
Check ventilator pressures.
Check the patient circuit.
Press autoset to for new alarm
thresholds.
Symptom Possible Cause Remedy