x = 1 exponential detector.
#7,SE,x;
Set detector for statistical analysis.
where:
x = 0 linear detector,
x = 1 exponential detector.
Response format:
#7,SE;
#7,SL;
Get statistical levels.
Response format:
#7,SL,x1,x2,x3,x4,x5,x6,x7,x8,x9,x10;
where:
x1..x10 statistical levels.
#7,SL,n,xn;
Set statistical level.
where:
n statistical level index (1..10),
xn statistical level of index n (1..99).
Response format:
#7,SL;
#7,LR;
Get Rolling Leq time values.
Response format:
#7,LR,x1[m],x2[m];
where:
x1 rolling time 1
st
value (1..59 if seconds, 1m..60m if minutes),
x2 rolling time 2
nd
value (1..59 if seconds, 1m..60m if minutes)
#7,LR,n,xn;
Set Rolling Leq time value.
where:
n result index (1 or 2),
xn rolling time n
th
value (1..60 if seconds, 1m..60m if minutes).
Response format:
#7,LR;
#7,FS;
Get file system version.
Response format:
#7,FS,x.xx;
where:
x.xx file system version number.