SVAN 977A User Manual - Appendixes
D.2. VIBRATION LEVEL METER
D.2.1 Basic terms and definitions (VLM mode)
T
current time period of the measurement in seconds.
exponential time constant in seconds for the giving time-weighting. The following time
constants are available: 100 ms, 125 ms, 200 ms, 500 ms, 1.0 s, 2.0 s, 5.0 s, 10.0 s.
W
frequency-weighting filter (HP, HP1, HP3, HP10, Vel1, Vel3, Vel10, VelMF, Dil1, Dil3,
Dil10, or Wh).
( )
ta
W
instantaneous frequency-weighted signal with the weighting filter W.
( )
ta
W
instantaneous frequency and time-weighted
signal with the weighting filter W and time
constant
calculated from the equation:
( ) ( )
( )
=
t
/t
dea
1
ta
2
WW
,
where:
variable of integration.
( )
tv
instantaneous vibration signal depends on
the <RMS Integration> parameter:
( )
( )
( )
=
=
=
Exp nIntegratioRMS
Lin nIntegratioRMS
ta
ta
tv
W
W
D.2.2 Definitions and formulas of the VLM result