Note: The RPM results are always registered in the logger file as logger results (with the logger
step) and as summary results (with the integration period step).
5.9 SETTING UP THE EXPOSURE TIME - EXPOSURE TIME
The Exposure Time parameter defines the
workday exposure time that is used for the
calculation of LEPd (cf. Appendix D for
definitions). This position is available only in
the sound mode. The Exposure Time values
are within the range [00h01, 08h00].
<ENT>
5.10 STATISTICAL LEVELS SETTINGS - STATISTICAL LEVELS
The Statistical Levels position is available
only in the Sound meter modes.
In the Statistical Levels screen, you can
define ten statistical levels, named from N1 to
N10, to be calculated, displayed and saved in
a file as Summary results.
<ENT>
The default statistical levels have following settings: 10, 20, 30, 40, 50, 60, 70, 80, 90 and 95. All values should
be within the integer range [1, 99]. Each individual value can be set independently from others.
5.11 PROGRAMMING THE INSTRUMENTS INTERNAL TIMER TIMER
The Timer function is used to programme the automatic start-up of the measurement at a given time and day
and with the parameters set in the Measurement section.
The Timer position allows you to programme
the internal real-time clock to act as a delayed
start and stop timer. The instrument will be
switched on automatically at the programmed
time and will perform the measurement with
the same settings used before the instrument
was turned off.
<ENT>
Modes of the timer function
The timer can be switched off (Off), switched
on only once (Single), or switched on many
times regularly (Multiple) with the period
between two consecutive measurements
(Repetition).
In the case the timer is active (Single or Multiple) and the instrument is switched on the clock icon appears
until finishing the programmed measurements.