161
SVAN 977A User Manual - Appendixes
#7,EP;
Get states of the External Power functionality.
Response format:
#7,EP,x,V
Off
,y,V
On
;
#7,EP,x,V
Off
,y,V
On
;
Set states of the External Power functionality.
Response format:
#7,EP;
where:
x = 0 Power Off functionality disabled,
x = 1 Power Off functionality enabled (the instrument will be switched-off after reaching
the threshold level V
Off
),
V
Off
switch-off threshold level in [V] (from 6.0 to 14.5),
y = 0 Power On functionality disabled,
y = 1 Power On functionality enabled (the instrument will be switched-on after reaching
the threshold level V
On
),
V
On
switch-on threshold in [V] (from 6.5 to 15.0).
note:
if (x = 1) and (y = 1) then (V
On
V
Off
) must be greater then or equal to 0.5V.
#7,BT;
Get state of the Bluetooth interface.
Response format:
#7,BT,x;
#7,BT,x;
Set state of the Bluetooth interface.
Response format:
#7,BT;
where:
x = 0 Bluetooth disabled,
x = 1 Bluetooth enabled.
#7,BR;
Get Bluetooth version.
Response format:
#7,BR,x.xx;
where:
x.xx Bluetooth version number.
#7,AC;
Get state of the Auto Calibration functionality.
Response format:
#7,AC,x;