Perform direct functions
July 29, 2013, 715003794 Rev. B 125
System Prep System Prep is a simple-to-use
alternative to running the
individual functions for Wet Prime,
Purge Injector, Equilibrate, and
Condition Column (described later
in this chapter). It performs a
series of automated steps to fully
ready the separations module for
running samples.
“Run System Prep”
on page 128
Wet Prime Replaces solvent in the path from
the reservoirs to V1 and V2 and
then to waste (V3). Use to change
the solvent(s) in the system.
“Wet prime the
system” on
page 129
Purge Injector Removes mobile phase from the
sample loop and syringe, replacing
it with fresh mobile phase. As an
option, performs a compression
test.
“Purge the injector”
on page 131.
Purge 2410/410
Reference
(if applicable)
Purges the Waters 2414, 2410, or
410 RI Detector reference cell.
“Purge the 2410
and 410 reference
cell” on page 132
Equilibrate Delivers mobile phase at the
current conditions for the specified
period of time. Equilibrates the
temperature in the sample
compartment and column heater
(or column heater/cooler) at preset
values.
“Equilibrate the
system” on
page 132
Condition
Column
Delivers mobile phase using the
gradient table specified in the
current separation method without
injecting a sample or running the
Events table.
“Condition the
column” on
page 133
Inject Samples Injects a sample one or more times
from the specified vial(s) using the
selected method.
“Inject samples” on
page 133
Direct functions: (Continued)
Direct function Description Reference