6 Create Methods, Sample Sets, and Sample Templates
172 July 29, 2013, 715003794 Rev. B
Edit the detector table
In the detector table, you program 2487 and 486 detector events to occur
during a run.
To enter events in the detector table:
1. Press the Table screen key in the Detectors screen. The following figure
shows the screen for a Waters 2487 detector using channels 1 and 2.
Temperature:
External Temp
°C 1
Sets the temperature of external
column heater (or column
heater/cooler) 1 in °C
1
.
Ambient to 150 °C
in 1 °C increments
Temperature:
External Temp
°C 2
Sets the temperature of external
column heater (or column
heater/cooler) 2 in °C
1
.
Ambient to 150 °C
in 1 °C increments
Autozero on
inject
Resets the output of the detector to 0
volts when the injection begins.
Selected or
deselected
1. The Waters 2414, 2410, and 410 Differential Refractometers can control the temperature of
one or two Waters Column Heater Modules (CHM) or column heater/cooler modules. These
devices are separate from the column heater or column heater/cooler attached to the
separations module.
2414, 2410, and 410 RI parameters: (Continued)
Parameter Function Value range