Create and edit a sample set
July 29, 2013, 715003794 Rev. B 175
3. Click the Sample Set screen key.
4. Name the sample set, and then click Enter.
Tip: The separations module automatically adds the appropriate
extension.
5. Enter information in the Sample Set table as appropriate.
6. Press Exit to display a dialog box that asks if you want to save the
changes.
The following table describes the parameters in the Sample Set table.
Sample Set table parameters:
Parameter Function Value range
Vials Selects the vial(s) from which the
injection is to be made. Use the “.”
key to specify a vial range. For
example, press 2.5 to enter vials 2
through 5.
1 to 120
Function The second column in the sample set
table specifies the function you apply
to a separation method. The
following table describes the
functions available with the
separations module. You use these
functions in the same manner as the
direct functions described in ““Direct
functions:” on page 124.
See the table titled
“Sample Set
functions:” on
page 176
Method Selects the separation method to use.
See “Create and edit separation
methods” on page 146 for the
procedures to create, edit, and store
separation methods.
Stored methods
Injs Number of injections to perform
using the current vial(s).
1 to 99