WM 68401 04/2021
MEDUCORE Standard
2
EN 133
6 Operation
Result The device carries out a cardiac rhythm analysis. The cardiac
rhythm analysis has one of two results:
Shock required (see “ Shock required”, page 133)
or
Shock not possible (see “ Shock not possible”, page 135)
Shock required
The device carries out a cardiac rhythm analysis, charges for shock
delivery and outputs the message:
If, based on the cardiac rhythm analysis, the device determines that
a shock is required, the device outputs the message:
The shock button flashes and an audible alarm sounds.
Voice prompt AED instruction
Stand clear of the patient Stand clear of the patient
Cardiac rhythm is being analyzed Analysis
Voice prompt AED instruction
Shock required Shock required
Press shock button Press shock button
Risk of injury from electric shock!
The electric shock administered to the patient may injure the user
or bystanders.
Stand clear of the patient.
Keep patient away from liquids (e.g. blood, gel or saline
solution).
Do not touch parts in contact with the patient (e.g. bed frames
or stretchers).
Keep your distance from liquids in contact with the patient.
Clearly warn bystanders to stand clear of the patient or parts in
contact with the patient and to keep their distance from liquids
in contact with the patient.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Weinmann MEDUCORE Standard2 and is the answer not in the manual?

Weinmann MEDUCORE Standard2 Specifications

General IconGeneral
BrandWeinmann
ModelMEDUCORE Standard2
CategoryMedical Equipment
LanguageEnglish

Related product manuals