EasyManua.ls Logo

YOKOGAWA UT55A - Environmental Conditions

YOKOGAWA UT55A
476 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
19-9
IM 05P01C31-01EN
Specifications
19
19.1HardwareSpecifications
19.1.16 EnvironmentalConditions
NormalOperatingConditions
Ambient temperature: -10 to 50ºC (-10 to 40ºC for side-by-side close mounting)
For the CC-Link option, 0 to 50 ºC (0 to 40 ºC for side-by-side close mounting)
• Ambienthumidity:20to90%RH(nocondensationallowed)
Magnetic field: 400 A/m or less
• Continuousvibrationat5to9Hz:Halfamplitudeof1.5mmorless,
1oct/min for 90
minutes each in the three axis directions
Continuousvibration
at9to150Hz:4.9m/s
2
or less, 1oct/min for 90 minutes each in
the three axis directions
Short-period vibration: 14.7 m/s
2
, 15 seconds or less
Shock: 98 m/s
2
or less, 11 ms
Altitude: 2000 m or less above sea level
Warm-up time: 30 minutes or more after the power is turned on
Startup time: Within 10 seconds
*: The LCD (a liquid crystal display) is used for a display portion of this product.
The LCD has a characteristic that the display action becomes late at the low temperature.
However, the control fuction is not affected.
Transportation and Storage Conditions
Temperature: -25 to 70ºC
Temperature change rate: 20ºC/h or less
• Humidity:5to95%RH(nocondensationallowed)
EffectsofOperatingConditions
Effect of ambient temperature:
VoltageorTCinput:±1µV/ºCor±0.01%ofF.S./ºC,whicheverislarger
Currentinput:±0.01%ofF.S./ºC
RTD input: ±0.05ºC/ºC (ambient temperature) or less
Analogoutput:±0.02%ofF.S./ºCorless
Effect of power supply voltage fluctuation
Analoginput:±0.05%ofF.S.orless
Analogoutput:±0.05%ofF.S.orless
(Each within rated voltage range)