892 J006 MNAH | PGC5000 G EN 2 | 57
Message
Issue / Fix
Too many script TCFs in
[Analysis Name]
The count of the Script TCFs in this Analysis exceeds the maximum limit.
Invalid min Method TCF
StartTime=%d < 0
A TCF at Method scope was found to have a time offset of less than zero
seconds.
Found more than one
Stream Step TCF.
Only one Stream Step is allowed in each Analysis.
More than one Stream Step
TCF in [AnalysisName]
No Analysis can have more than one Stream Step.
Missing Stream Step TCF
Every Analysis must have a Stream Step.
NO TCFs in Analysis:
"[AnalysisName]"
Analysis did not contain any TCFs.
[Method Name] inside prev
Method
An Analysis was found containing a Method having no TCFs nor any
Sequences.
[Sequence Name] in prev
Sequence
An Analysis was found containing a Sequence having no TCFs.
All active STEP table items
have a bad ["Analysis"]
No Schedule Elements in the STEP table were found having a well-
formed Analysis.
[Component Name] - bad
time values
A part of the named Component TCF is outside of its containing Method.
Malformed Method -> no
oven assignment
Every Method must have an Oven Name specified in that Method.
Sequence.TimeOffset >=
Method.CycleTime
No Sequence start time can be greater than the containing Method's
CycleTime.
SkipNextIf is the last TCF in
Analysis
A Skip Next TCF type cannot be the last TCF in an Analysis.
SkipNextIf TCF precedes
Component TCF
A Skip Next If TCF cannot precede Component TCF.
SkipNextIf TCF precedes
DoNextIf TCF
A Do Skip If TCF cannot precede another Do Skip If TCF.
SkipNextIf TCF precedes
TcfSkipNextIf
A Skip Next If TCF cannot precede a Do Next If TCF.
SkipNextIf TCF precedes
UnknownPeak TCF
A Skip Next If TCF cannot precedes an Unknown Peak TCF type.
Malformed Analysis: Startup
purge < 1
An Analysis's Startup Purge must be greater than zero.
[Method Name]:Skip then
StreamStep
In the named Method, a skip occurs before the Stream Step.
SkipNextIf TCF precedes
StreamStep Tcf
A Skip Next If TCF cannot precede a Stream Step TCF.
StreamStep TCF starts at
time < 1 second
A Stream Step TCF cannot start at a time of less than one second into
the CycleTime.
(MaxTcfEndTime = [Secs])
> (CycleTime= [Secs])
Some TCF in the Method exceeded this containing Method's CycleTime.
Not all TCFs have a valid
start time
A TCF's time offset is configured at less than one second before the end
of the Cycle Time. Note one exception: End-Of-Analysis scripts.
1st TCF not starting at
reasonable time
First TCF time offset is configured to start before one second into the
Cycle Time.
Invalid min Seq TCF
StartTime = %d < 0
A TCF at Sequence scope has a time offset of less than zero seconds.
[Description] TCF StartTime
<= 0
No TCFs are allowed to start at a start time of <= '0' (Inside a Method /
Sequence).
Same oven address in 2
Methods found Analysis:
"[AnalysisName]"
In one Analysis, two methods are configured to run in the same oven.
Same Component in two
different Methods
The same Component Name is found in two different Methods.