94 | PGC5000 GEN 2 | 892 J006 MNAH
5.5.5 Do next if
This allows execution of the next TCF in the method only if the stream value equals the stream name
entered in the method. This function allows adding a specific TCF to a particular stream, but not to
every stream.
Table 5-7: Do next if
Parameter
Range
Units
Description
User-definable
Text
Time Offset
0 to 14400
Seconds
Stream
Available from configuration
N/A
5.5.6 Peak threshold
Table 5-8: Peak threshold
Parameter
Range
Units
Description
User-definable
Text
Time Offset
0 to 14400
Seconds
Threshold
1 to 99 (default 6)
Numeric
Detector Channel
Select from configuration
N/A
5.5.7 Pressure check
Pressure check, which is part of the Electronic Pressure Controller, allows for pressure zone variances.
A pressure value outside the pressure limit range causes an alarm.
Table 5-9: Pressure check
Parameter
Range
Units
Description
User-definable
Text
Time Offset
0 to 14400
Seconds
Zone
Available from configuration
N/A
Action
Ignore, Report, Skip, Abort
N/A
Low Limit
0 to 100
PSI
High Limit
0 to 100
PSI
5.5.8 Pressure control
This feature of the Electronic Pressure Controller allows review or modification of the pressure zone
set point and ramp rate.
Table 5-10: Pressure control
Parameter
Range
Units
Description
User-definable
Text
Time Offset
0 to 14400
Seconds
Zone
Available from configuration
N/A
Setpoint
0 to 100
PSI
Ramp Rate
0 to 1000
PSI/min
5.5.9 Pressure default
Returns zone to Setup Tab isobaric setpoint.
Table 5-11: Pressure default
Parameter
Range
Units
Description
User-definable
Text
Time Offset
0 to 14400
Seconds
Zone
Available from configuration
N/A