892 J006 MNAH | PGC5000 G EN 2 | 117
temperature feedback. Check for
other faults.
System information does not
match DTC internal setting.
Air or carrier pressure switch is
open.
Verify proper air/carrier pressure and
increase if necessary. If pressure is
correct, replace switch.
DTC unable to read
temperatures.
Replace temperature probe. Possibly
replace DTC Digital. Refer to the
Service Manual.
Device did not communicate
with OC or go online.
Check board LEDs for error. Refer to
the Service Manual.
Device stopped
communicating.
Check board LEDs for error. Check
CAN bus.
Refer to the Service Manual.
Failed to send executable code
to device.
Check board LEDs for error. Refer to
the Service Manual.
Device did not communicate
with OC or go online.
Check board LEDs for error.
Device stopped
communicating.
Check board LEDs for error. Check
CAN bus.
Refer to the Service Manual.
Failed to send executable code
to device.
Check board LEDs for error. Refer to
the Service Manual.
RTC signal missing or
irregular.
Check CAN connections. Check power
and ground connections. Check
chassis ground connection to Oven
Electronics Door.
7 Scripting
7.1 Introduction
The PGC5000 Generation 2 analyzer has programmatic control and calculation in addition to the built-
in capabilities. The Program Tab provides the means to customize analyzer operation to support most
requirements (see Figure 7-1).