892 J006 MNAH | PGC5000 G EN 2 | 117
Source
Group
Issue
Resolution
temperature feedback. Check for
other faults.
T-Rating mismatch
DTC#.
IsoThrml Oven
System information does not
match DTC internal setting.
Contact ABB.
Low Prsr-Air/Carrier
DTC#.
IsoThrml Oven
Air or carrier pressure switch is
open.
Verify proper air/carrier pressure and
increase if necessary. If pressure is
correct, replace switch.
Temp Reading Invalid
DTC#.
IsoThrml Oven
DTC unable to read
temperatures.
Replace temperature probe. Possibly
replace DTC Digital. Refer to the
Service Manual.
CAN Node Not Found
EPC
Device did not communicate
with OC or go online.
Check board LEDs for error. Refer to
the Service Manual.
Lost CAN Heartbeat
EPC
Device stopped
communicating.
Check board LEDs for error. Check
CAN bus.
Refer to the Service Manual.
Code Download Error
EPC
Failed to send executable code
to device.
Check board LEDs for error. Refer to
the Service Manual.
Barometric pressure
low
EPC
Status Only.
Not Applicable.
Barometric pressure
high
EPC
Status Only.
Not Applicable.
Ambient temperature
low
EPC
Status Only.
Not Applicable.
Ambient temperature
high
EPC
Status Only.
Not Applicable.
Zone plateau
EPC.Carrier#
Status Only.
Not Applicable.
Zone In Band
EPC.Carrier#
Status Only.
Not Applicable.
Zone Fail Zero
EPC.Carrier#
Status Only.
Not Applicable.
Zone Enabled
EPC.Carrier#
Status Only.
Not Applicable.
Can Node Not Found
Ovn.DetAmp#
Device did not communicate
with OC or go online.
Check board LEDs for error.
Lost CAN Heartbeat
Ovn.DetAmp#
Device stopped
communicating.
Check board LEDs for error. Check
CAN bus.
Refer to the Service Manual.
Code Download Error
Ovn.DetAmp#
Failed to send executable code
to device.
Check board LEDs for error. Refer to
the Service Manual.
RTC Failure
Ovn.DetAmp#
RTC signal missing or
irregular.
Check CAN connections. Check power
and ground connections. Check
chassis ground connection to Oven
Electronics Door.
7 Scripting
7.1 Introduction
The PGC5000 Generation 2 analyzer has programmatic control and calculation in addition to the built-
in capabilities. The Program Tab provides the means to customize analyzer operation to support most
requirements (see Figure 7-1).