78 | PGC5000 GEN 2 | 892 J006 MNAH
Figure 4-49: Subscriber Added and Activated example
The following table identifies all events which trigger output to the defined subscriber lists.
Table 4-6: Subscriber descriptions
Event
Trigger/Origination Point
Associated Indicator
Power Recovery: <DateTime>
Analyzer completes power up sequence
Report Only
Power Fail: <DateTime>
Detects of loss of power
MC: PowerFail
User Login: <Username>
User logs in to the analyzer
Report Only
User Logout: <Username>
User logs out of the analyzer
Report Only
Stream Online:
<StreamName>
Stream set online manually, programmatically
(via scripting) or automatically (via DCS call
through OPC Server)
StreamActive
Stream Offline:
<StreamName>
Stream set offline manually, programmatically
or automatically
StreamActive
Schedule Run:
<ScheduleName>
Resource Group has been activated manually,
programmatically or automatically
Report Only
Schedule Stopped:
<ScheduleName>
Resource Group has been deactivated manually,
programmatically or automatically
Report Only
Schedule Paused:
<ScheduleName>
Resource Group suspended manually,
programmatically or automatically
Report Only
On-Line: <ScheduleName>
Schedule runs Analyses
OnLine
Idle: <ScheduleName>
Schedule has no Analyses to Run
OnLine
Calibration Started:
<ScheduleName>
Schedule starts calibration
Calibrating
Calibration Competed:
<ScheduleName>
Schedule completes calibration
Calibrating
Validation Started:
<ScheduleName>
Schedule starts validation
Validating
Validation Completed:
<ScheduleName>
Schedule completes a validation
Validating
Maintenance Started:
<ScheduleName>
Schedule Maintenance Starts: Maintenance
Button Activated
Maintenance
Maintenance Completed:
<ScheduleName>
Schedule Maintenance Ends: Maintenance
Button deactivated
Maintenance
The subscriber list supports failover (redundancy). If the secondary network interfaces are configured,
the system switches to the secondary network and continues to report data if the primary network
fails (activates COM failure indicator).