START UP and SHUTDOWN
AEROSET
®
Customer Training Guide 3-11
94858-103 — July 2003
When the System is configured to Save by Carrier
ID (no sample bar codes) Sample Carrier/Positions
cannot be reused and must be cleared using this
procedure.
The Delete function on the Database screen will
not clear the Database.
NOTE: If the Cal, QC option or Cal, QC, QC Avg option
is selected instead of the All Records option to Clear
Database, then:
C/P, SID, and PID are deleted for complete,
incomplete, and validated results; these
samples become preserved samples.
A search cannot be performed for a preserved
sample.
Text color on the DATABASE screen is green
for a preserved sample.
The C/P, SID, and PID of the preserved samples
can be used for new patient samples.
<Rxn Graph> cannot be selected to view a
reaction graph for a preserved sample.
NOTE: If the Wash Cuvettes option is not performed as
part of the SHUTDOWN Procedure, Cuvette Integrity
Check data is not reset. Refer to Section 10,
Troubleshooting and Diagnostics for more information.