MODULE 3: DAILY START UP AND CHECKS
3-12 AEROSET
®
Customer Training Guide
94858-103 — July 2003
Performing the START UP Procedure
1. Select <START UP> in the Action Area to display the START UP
OPTIONS screen.
Figure 3.3: START UP OPTIONS Screen
2. Place a checkmark (;) in the box beside each desired option by
touching the box.
3. Ensure there are sufficient volumes of all solutions required.
NOTE: Refer to Solutions Used for START UP,
SHUTDOWN, and Daily Operation and Solution
Volume Display within this Module for details.
4. Select <Start> to begin the START UP Procedure.
NOTE: If the power to the System is turned OFF, when
power is reinitiated, the text on the START UP button is
pink. The text turns black if all procedures complete
successfully.
NOTE: A pink line below the text of any START UP
or SHUTDOWN option indicates the option was
not completed successfully, or power to the System
was turned off. Review any Error Log Message that
was generated, then repeat the START UP or
SHUTDOWN option that was underlined.
Can use screens and order samples
during START UP.
Show how to setup START UP and
SHUTDOWN screens.