MODULE 6: CALIBRATION
6-24 AEROSET
®
Customer Training Guide
94858-103 — July 2003
Viewing Stored Calibration Curves
Calibration data can be stored in the Stored Calibration Summary file. If
the calibration data is not saved to the Stored Calibration Summary, it is
not available for display on the Assay Status screen.
NOTE: For information about saving calibration data to the
Stored Calibration Summary, refer to the CALIBRATION
SUMMARY screen section in this Module.
To view the stored calibration curves, perform the following steps:
1. Select on the lower right-hand corner of the Calibration
Information area of the ASSAY STATUS screen.
2. The absorbance data and curve from the previous calibration curve
are displayed.
3. Select .
4. The absorbance data and curve of the most recent stored calibration
curve are displayed. The date and time the stored calibration was
performed are displayed in the lower right-hand corner of the
Calibration Information area. The stored calibration curve and the
date and time performed are pink.
5. Continue to select to scroll through all stored calibration curves
(up to 31 can be stored) from the most recent to the earliest
performed.
6. Select to scroll back through to the most recent calibration, or
select to return to the current calibration data.