Calibration Review
AEROSET
®
Customer Training Guide 6-31
94858-103 — July 2003
Deleting the Stored Average Data
To delete the stored average data for an assay, perform the following
procedure:
NOTE: The System status must be READY to perform this
procedure.
1. Select <Cal Summary> in the Information Access Area. The
CALIBRATION SUMMARY screen is displayed.
2. Select <Stored> in the right hand column of the CALIBRATION
SUMMARY screen. The stored averages are displayed.
3. Select the assay for which the stored average data are to be deleted
by selecting the assay name.
4. Select <Delete> in the right-hand column. A confirmation dialog
window is displayed.
Figure 6.14: Delete Confirmation Dialog Window
5. Select <OK>. The stored average calibration data are deleted.