Reviewing QC Results
AEROSET
®
Customer Training Guide 7-19
94858-103 — July 2003
3. The following information is displayed on the Details dialog
window:
NOTE: The mean, SD, and CV are calculated using all
data points except those excluded in the Levey-Jennings
Details screen.
4. To exclude the displayed value, select the Exclude option at the
bottom of the Details dialog window.
NOTE: If a date range is configured for the assay:
QC values for the selected date range are blue.
Excluded QC values are white.
QC values outside the date range are gray.
Item Description
Value
The result value of the selected QC point
NOTE: An “L” or “H” Flag is displayed
next to the result if the value is outside the
defined 2 SD range. A Result Error Code is
displayed if an error occurred for the result.
Min
The lowest result value in the QC data
Max
The highest result value in the QC data
Date
The completion date and time of the selected QC
point
Mean
The actual mean of the QC data
SD
The actual standard deviation of the QC data
CV
The actual coefficient of variation of the QC data
Operator
The User Code of the Operator logged onto the
AEROSET System when the selected QC value was
completed
Include/Exclude
The selected QC value can be included or excluded
in the calculation of the QC statistics. Excluded
values are not deleted.
Comment
A comment can be selected from the drop-down list
box or added for a QC value
Excluded by
Date Range
Message displays with pink text when a QC value is
outside the selected QC date range