MODULE 1: SYSTEM TOUR
1-22 AEROSET
®
Customer Training Guide
94858-103 — July 2003
ICT™ Cleaning Fluid
The ICT Cleaning Fluid (deproteinizer) is placed in the Calibrator/
Control Carousel and is used daily during the START UP or
SHUTDOWN Procedures to clean the ICT Module. The lyophilized
material is reconstituted with the supplied diluent. Unopened bottles are
stored at room temperature. Once reconstituted (opened), the bottles
must be refrigerated.
Figure 1.14: ICT Cleaning Fluid
ICT Reference Solution
The ICT Reference Solution container is loaded onto a weight platform
behind the front-right door. It is aspirated and analyzed by the ICT Unit,
before and after each sample, to provide a reference concentration used
to calculate results.
Wash Solutions
CAUTION: Chemical Hazard. Refer to Appendix,
Hazards of this manual.
There are two Wash Solutions used on the AEROSET System:
Alkaline Wash Solution
Acid Wash Solution
The containers are loaded onto a weight platform behind the front-right
door. The Wash Solutions are used by the Cuvette Washer to clean the
cuvettes after sample analysis.
Good 14 days after opening
at 2-8°C.
125640
Unopened - Store at room
temperature.
!
Caustic - need gloves and
goggles.
½ Liter
Unopened - Store at room
temperature.