Checking Reagent Inventory
AEROSET
®
Customer Training Guide 2-11
94858-103 — July 2003
Load Reagents
Rules for loading reagent cartridges:
R1 reagents must always reside in Reagent Supply Center 1.
R2 reagents must always reside in Reagent Supply Center 2.
AEROSET System assays are divided into two distinct categories
(A-Line and B-Line) to allow the System to process tests at the
fastest possible throughput.
A-Line Tests—reagents always reside in segments A, B,
or C.
B-Line Tests—reagents always reside in segment D
New cartridges should always be loaded in a higher numbered
position than cartridges of the same reagent already onboard.
The System always aspirates first from the cartridge in the lowest
numbered position.
For assays using two reagent cartridges, the R1/R2 pair must be
placed onboard at the same time. Refer to the assay-specific
Package Inserts for specific reagent handling instructions.
After loading reagents, a reagent scan must be performed to update
the inventory