EasyManuals Logo
Home>Agilent Technologies>Laboratory Equipment>1220 Infinity II LC

Agilent Technologies 1220 Infinity II LC User Manual

Agilent Technologies 1220 Infinity II LC
398 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #143 background imageLoading...
Page #143 background image
1220 Infinity II LC System User Manual 143
Test Functions and Calibration
9
1220 Infinity II LC System
1220 Infinity II LC System
Installation Check
The Installation Check switches on all available modules, purges the system for
five minutes at 1 mL/min, tests the flow path by applying a pressure up to
200 bar and switches on the oven (if available) and detector.
The Installation Check passes if the following conditions are met:
• All modules switch on successfully within the timeout period (120 s).
• The pump achieves 200 bar after 5 min.
• The oven reaches 2 K above its actual temperature.
• The detector lamp ignites and the detector reaches a Ready state.
NOTE
The pump and detector are mandatory for this check; the oven and autosampler are
optional.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1220 Infinity II LC and is the answer not in the manual?

Agilent Technologies 1220 Infinity II LC Specifications

General IconGeneral
BrandAgilent Technologies
Model1220 Infinity II LC
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals