EasyManua.ls Logo

Agilent Technologies 1220 Infinity II LC

Agilent Technologies 1220 Infinity II LC
398 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1220 Infinity II LC System User Manual 231
Error Information
10
VWD Detector Error Messages
Grating or Filter Motor Errors
Error ID: Grating: 7800, 7801, 7802, 7803, 7804, 7805, 7806, 7808, 7809;
Filter: 7810, 7811, 7812, 7813, 7814, 7815, 7816
The motor test has failed.
During the motor tests, the detector moves the motor to the end position while
monitoring the end-position sensor. If the end position is not found, the error
message is generated.
Probable cause Suggested actions
1
Filter motor is not connected.
Please contact your Agilent service
representative.
2
Cable/connector defective.
Please contact your Agilent service
representative.
Test 0 F ailed : Filter motor.
Test 1 F ailed : Grating motor.
Probable cause Suggested actions
1
Motor is not connected.
Please contact your Agilent service
representative.
2
Defective motor.
Please contact your Agilent service
representative.
3
Defective/missing grating or filter.
Please contact your Agilent service
representative.
4
Cable/connector defective.
Please contact your Agilent service
representative.

Table of Contents