236 1220 Infinity II LC System User Manual
10
Error Information
DAD Detector Error Messages
Diode Current Leakage
Error ID: 1041
When the detector is switched on, the processor checks the leakage current of
each of the optical diodes. If the leakage current exceeds the upper limit, the
error message is generated.
UV Ignition Failed
Error ID: 7452
The UV lamp failed to ignite.
The processor monitors the UV lamp current during the ignition cycle. If the
lamp current does not rise above the lower limit within 2 – 5 seconds, the
error message is generated.
Probable cause Suggested actions
1
Defective connector or cable.
Please contact your Agilent service
representative.
2
Defective power supply.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Defective PDA/optical unit.
Please contact your Agilent service
representative.
2
Defective connector or cable.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1220 Infinity II LC and is the answer not in the manual?

Agilent Technologies 1220 Infinity II LC Specifications

General IconGeneral
BrandAgilent Technologies
Model1220 Infinity II LC
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals