1220 Infinity II LC System User Manual 95
Solvent Delivery System Description
5
Variable Stroke Volume
Variable Stroke Volume
Due to the compression of the pump-chamber volume, each plunger stroke of
the pump generates a small pressure pulsation, influencing the flow ripple of
the pump. The amplitude of the pressure pulsation is dependent mainly on the
stroke volume and the compressibility compensation for the solvent in use.
Small stroke volumes generate pressure pulsations of smaller amplitude than
larger stroke volumes at the same flow rate. In addition, the frequency of the
pressure pulsations are higher. This decreases the influence of flow pulsations
on quantitative results.
In gradient mode, smaller stroke volumes result in less flow ripple and
improve composition ripple.
The pump uses a processor-controlled spindle system to drive its plungers.
The normal stroke volume is optimized for the selected flow rate. Low flow
rates use a small stroke volume, while higher flow rates use a larger stroke
volume.
When the stroke volume for the pump is set to AUTO mode, the stroke is
optimized for the flow rate in use. A change to larger stroke volumes is
possible but not recommended.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1220 Infinity II LC and is the answer not in the manual?

Agilent Technologies 1220 Infinity II LC Specifications

General IconGeneral
BrandAgilent Technologies
Model1220 Infinity II LC
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals