18 1220 Infinity II LC System User Manual
2
Site Requirements and Specifications
Site Requirements
Environment
Your Agilent 1220 Infinity II LC will work within specifications at ambient
temperatures and relative humidity as described in the following sections.
ASTM drift tests require a temperature change below 2 °C/hour (3.6 °F/hour)
measured over one hour period. Our published drift specification is based on
these conditions. Larger ambient temperature changes will result in larger
drift.
Better drift performance depends on better control of the temperature
fluctuations. To realize the highest performance, minimize the frequency and
the amplitude of the temperature changes to below 1 °C/hour (1.8 °F/hour).
Turbulences around one minute or less can be ignored.
Condensation within the module
Condensation can damage the system electronics.
âž” Do not store, ship or use your module under conditions where temperature
fluctuations could cause condensation within the module.
âž” If your module was shipped in cold weather, leave it in its box and allow it to warm
slowly to room temperature to avoid condensation.