1220 Infinity II LC System User Manual 217
Error Information
10
Autosampler Error Messages
Initialization Failed
Error ID: 4020
The autosampler failed to complete initialization correctly.
The autosampler initialization procedure moves the needle arm and transport
assembly to their home positions in a predefined sequence. During
initialization, the processor monitors the position sensors and motor encoders
to check for correct movement. If one or more of the movements is not
successful, or is not detected, the error message is generated.
Initialization with Vial
Error ID: 4028
4
Defective sample transport assembly flex
board.
Please contact your Agilent service
representative.
5
Defective main board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
Probable cause Suggested actions
1
Mechanical obstruction.
Ensure unobstructed movement of the
transport assembly.
2
Defective sampling unit flex board.
Please contact your Agilent service
representative.
3
Defective transport assembly flex board.
Please contact your Agilent service
representative.
4
Defective sampling unit motor.
Please contact your Agilent service
representative.
5
Defective main board.
Please contact your Agilent service
representative.