EasyManua.ls Logo

Agilent Technologies 1220 Infinity II LC

Agilent Technologies 1220 Infinity II LC
398 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
1220 Infinity II LC System User Manual 301
Maintenance and Repair
11
Variable Wavelength Detector (VWD)
Correcting Leaks
1 Remove the lower front cover.
2 Use tissue to dry the leak sensor area.
3 Observe the capillary connections and the flow cell area for leaks and
correct, if required.
4 Replace the front cover.
When If a leakage has occurred in the flow cell area or at the capillary connections
Tools required Description
Tissue
Wrench, 1/4 inch
for capillary connections
Parts required Description
None
/HDNSODQH
:DVWHRXWOHW
/HDNVHQVRUDVVHPEO\

Table of Contents

Related product manuals