1220 Infinity II LC System User Manual 301
Maintenance and Repair
11
Variable Wavelength Detector (VWD)
Correcting Leaks
1 Remove the lower front cover.
2 Use tissue to dry the leak sensor area.
3 Observe the capillary connections and the flow cell area for leaks and
correct, if required.
4 Replace the front cover.
When If a leakage has occurred in the flow cell area or at the capillary connections
Tools required Description
Tissue
Wrench, 1/4 inch
for capillary connections
Parts required Description
None
/HDNSODQH
:DVWHRXWOHW
/HDNVHQVRUDVVHPEO\