1220 Infinity II LC System User Manual 197
Error Information
10
General Error Messages
General Error Messages
General error messages are generic to all Agilent series HPLC modules and
may show up on other modules as well.
Compensation Sensor Open
Error ID: 0081
The ambient-compensation sensor (NTC) on the power switch board in the
module has failed (open circuit).
The resistance across the temperature compensation sensor (NTC) on the
power switch board is dependent on ambient temperature. The change in
resistance is used by the leak circuit to compensate for ambient temperature
changes. If the resistance across the sensor increases above the upper limit,
the error message is generated.
Compensation Sensor Short
Error ID: 0080
The ambient-compensation sensor (NTC) on the power switch board in the
module has failed (open circuit).
Probable cause Suggested actions
1
Loose connection between the power
switch board and the main board
Please contact your Agilent service
representative.
2
Defective power switch assembly
Please contact your Agilent service
representative.
3
Defective main board.
Please contact your Agilent service
representative.