220 1220 Infinity II LC System User Manual
10
Error Information
Autosampler Error Messages
Missing Wash Vial
Error ID: 4035, 4542, 4707
The wash vial programmed in the method was not found.
When the gripper arm picks a vial out of the sample tray, the processor
monitors the gripper motor encoder. If a vial is present, the closing of the
gripper fingers is limited by the vial. However, if no vial is present, the gripper
fingers close too far. This is sensed by the processor (encoder position),
causing the error message to be generated.
Motor Temperature
Error ID: 4027, 4040, 4261, 4451
One of the motors of the transport assembly has drawn excessive current,
causing the motor to become too hot. The processor has switched OFF the
motor to prevent damage to the motor.
See figure Figure 35 on page 109 for motor identification.
• Motor 0 temperature: X-axis motor.
Motor 1 temperature: Z-axis motor.
Motor 2 temperature: Theta (gripper rotation) motor.
Motor 3 temperature: Gripper motor (motor for gripper fingers).
3
Defective gripper assembly (defective
gripper fingers or belt).
Exchange the gripper assembly.
4
Defective transport assembly flex board.
Please contact your Agilent service
representative.
Probable cause Suggested actions
Probable cause Suggested actions
1
No wash vial in the position defined in the
method.
Install the wash vial in the correct position, or
edit the method accordingly.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1220 Infinity II LC and is the answer not in the manual?

Agilent Technologies 1220 Infinity II LC Specifications

General IconGeneral
BrandAgilent Technologies
Model1220 Infinity II LC
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals