238 1220 Infinity II LC System User Manual
10
Error Information
DAD Detector Error Messages
Calibration Values Invalid
Error ID: 1036
The calibration values read from the spectrometer ROM are invalid.
After recalibration, the calibration values are stored in ROM. The processor
periodically checks if the calibration data are valid. If the data are invalid or
cannot be read from the spectrometer ROM, the error message is generated.
Holmium Oxide Test Failed
Probable cause Suggested actions
1
Defective connector or cable.
Please contact your Agilent service
representative.
2
Defective PDA/optical unit.
Please contact your Agilent service
representative.
Probable cause Suggested actions
1
Lamps switched off.
Ensure the lamps are switched on.
2
Defective or dirty flow cell.
Ensure the flow cell is inserted correctly, and is
free from contamination (cell windows, buffers
etc.).
3
Defective filter assembly.
Please contact your Agilent service
representative.
4
Defective achromat assembly.
Please contact your Agilent service
representative.
5
Defective PDA/optical unit.
Please contact your Agilent service
representative.