DAD Detector Error Messages
DAD Detector Error Messages
These errors are specific to the diode array detector.
Visible Lamp Current
The visible lamp current is missing.
The processor continually monitors the lamp current during operation. If the
current falls below the lower current limit, the error message is generated.
Visible Lamp Voltage
The visible lamp voltage is missing.
The processor continually monitors the voltage across the lamp during
operation. If the lamp voltage falls below the lower limit, the error message is
generated.
Probable cause Suggested actions
1
Lamp disconnected.
Ensure the visible lamp connector is seated
firmly.
2
Defective visible lamp.
Exchange the visible lamp.
3
Defective connector or cable.
Please contact your Agilent service
representative.
4
Defective power supply.
Please contact your Agilent service
representative.