EasyManuals Logo

Agilent Technologies 1220 Infinity II LC User Manual

Agilent Technologies 1220 Infinity II LC
398 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #213 background imageLoading...
Page #213 background image
1220 Infinity II LC System User Manual 213
Error Information
10
Pump Error Messages
MCGV Fuse
Error ID: 2043
Valve Fuse 0: Channels A and B
Valve Fuse 1: Channels C and D
The gradient valve in the quaternary pump has drawn excessive current
causing the electronic fuse to open.
Wait Timeout
Error ID: 2053
Probable cause Suggested actions
1
Gradient valve disconnected.
Ensure the gradient valve is connected
correctly.
2
Connection cable (inside instrument) not
connected.
Please contact your Agilent service
representative.
3
Connection cable (inside instrument)
defective.
Please contact your Agilent service
representative.
4
Gradient valve defective.
Exchange the gradient valve.
Probable cause Suggested actions
1
Defective gradient valve.
Restart the quaternary pump. If the error
message appears again, exchange the gradient
valve.
2
Defective connection cable (front panel to
main board).
Please contact your Agilent service
representative.
3
Defective main board.
Please contact your Agilent service
representative.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Agilent Technologies 1220 Infinity II LC and is the answer not in the manual?

Agilent Technologies 1220 Infinity II LC Specifications

General IconGeneral
BrandAgilent Technologies
Model1220 Infinity II LC
CategoryLaboratory Equipment
LanguageEnglish

Related product manuals